Currently under review by the US Food and Drug Administration (FDA), the Dexcom G7 is expected to be an improvement over the Dexcom G6 version in several ways.

The on-body size will be 60% smaller, "roughly the size of three stacked quarters," according to the authors, and will incorporate the sensor with a single-use transmitter, as opposed to the current separate 3-month transmitter used with the G6 sensor. This will eliminate the need for using a transmitter across multiple sensor sessions (as is the case for G6).

The warm-up period after insertion is reduced from 2 hours to 27 minutes, and users are given an extra 12-hour "grace period" after the 10-day wear period to change the device before it stops displaying glucose data. Up to 24 hours of missed data can also be recaptured.

"The enhanced features of G7 may increase clinical adoption, encourage sustained use, and reduce the burden of diabetes management," write Satish K. Garg, MD, of the University of Colorado, Aurora, and colleagues in their article, published online February 14 in Diabetes Technology & Therapeutics.  

Several features of the G6 remain unchanged, including factory calibration, but also the capacity for optional user calibrations, use of Bluetooth to transmit data up to 20 feet (approximately 6 meters), and data displays on either a dedicated receiver or a variety of iOS and Android smart devices.

It will also allow for user-customized settings and alerts, and the option for real-time remote "share" monitoring with caregivers or providers. The G7 will also not be susceptible to interference by acetaminophen (paracetamol) or ascorbic acid.

And, like the G6, the G7 was submitted for approval to the FDA as an "integrated CGM," meaning that it will be interoperable with other compatible devices including insulin pumps, glucose meters, or other electronic devices used for diabetes management.

Device Accuracy Demonstrated on Both Abdomen and Arm 

The prospective, multicenter, single-arm study reported by Garg and colleagues was conducted at 12 US sites between February and June 2021.

In-clinic visits were conducted on days 1 or 2, 4 or 7, and one additional day for comparisons with a reference glucose measure (YSI 2300 Stat Plus glucose analyzer). Participants wore blinded G7 sensors concurrently on the upper arm and abdomen while continuing to use their own personal glucose monitoring method (CGM or fingerstick) at home.

A total of 316 insulin-using adults with type 1 or type 2 diabetes contributed data from 308 arm- and 311 abdomen-placed blinded devices, which generated 77,774 matched pairs of data within the blood glucose reportable range of 40-400 mg/d (4 mmol/l = 72 mg/d)

The overall mean absolute relative difference (MARD) of each CGM-YSI pair (a standard metric for CGM analysis) was 8.2%, with 9.1% for the abdomen and 8.2% for the arm.

Accuracy remained high in both arm- and abdomen-placed sensors across the 10-day wear period through the 12-hour grace period and across glucose ranges. There were no significant differences between G7 accuracy by diabetes type, insulin regimen, or body mass index.

The highest agreement rates and lowest MARDs occurred when CGM readings were increasing or decreasing by no more than 1 mg/dL per minute. However, even at the highest rates of glucose concentration change, MARD values below 10% were found for arm-placed sensors and below 10.5% for abdomen-placed sensors, Garg and colleagues report.

When the hypoglycemia threshold alert was set to 55 mg/d (3 mmol/l), true alert rates for detection of hypoglycemia below 70 mg/dL by sensors worn on the arm and abdomen were 91.3% and 85.2%, respectively.

With hyperglycemia threshold alerts set to 300 mg/dL, the true alert rates for detection of hyperglycemia greater than 250 mg/dL by sensors worn on the arm and abdomen were 99.9% and 99.8%, respectively.

The overall mean time lag for the sensors was 3.5 minutes, 3.6 minutes for the arm and 3.4 minutes for the abdomen. There were no serious adverse events during the study.

The study excluded children and adolescents; data from these populations will be reported separately, the authors note.

Accuracy at Least as Good as Prior Dexcom Versions, Competitors

The MARD values of 8.2% on the arm and 9.1% on the abdomen were similar to or better than accuracy measurements of other commercially available CGM systems, note Garg and colleagues, although they acknowledge that few head-to-head studies at different anatomic locations have been conducted.

A study of an older Dexcom version (G4 Platinum) found MARD estimates of 12.0% on the arm and 12.3% on the abdomen, the authors note.  

The newly FDA-approved implantable Eversense E3 (Senseonics) CGM, worn on the upper arm, has a MARD of 9.1%, while the arm-placed Abbott FreeStyle Libre 2, approved in the US in June 2020, has an overall MARD of 9.3%.

Lag-time differences between the reference and G7 were also similar to or better than time delays in prior-generation Dexcom CGMs, Garg and colleagues say.

Participants also completed a survey. "The redesigned applicator allows for sensor deployment with one hand, and most participants found G7 easier to insert than their prior CGM system," the researchers say.

Finally, "unlike G6, G7 allows for temporary silencing of all audible alerts, including Urgent Low. Taken together, these attributes are anticipated to provide for a better end-user experience with G7 and help reduce diabetes burden," they conclude.

From www.medscape.com

 

Nyhetsinfo

Diabetes Technol Ther. Published online February 14, 2022. Abstract

https://www.liebertpub.com/doi/10.1089/dia.2022.0011

Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes

Abstract

Background:

We evaluated the accuracy and safety of a seventh-generation (G7) Dexcom continuous glucose monitoring (CGM) during 10 and a half days of use in adults with diabetes.

Methods:

Adults with either type 1 or type 2 diabetes (on intensive insulin therapy or not) participated at 12 investigational sites in the United States. In-clinic visits were conducted on Days 1 or 2, 4 or 7, and the second half of 10 or the first half of day 11 for frequent comparisons with comparator blood glucose measurements obtained with the YSI 2300 Stat Plus glucose analyzer. Participants wore sensors concurrently on the upper arm and abdomen. Accuracy evaluation included the proportion of CGM values within 15% of comparator glucose levels >100 mg/dL or within 15 mg/dL of comparator levels ≤100 mg/dL (%15/15), along with the %20/20 and %30/30 agreement rates.

The mean absolute relative difference (MARD) between temporally-matched CGM and comparator values were also calculated. Results: Data from 316 participants (619 sensors, 77,774 matched pairs) were analyzed. For arm- and abdomen-placed sensors, overall MARDs were 8.2% and 9.1%, respectively. Overall %15/15, %20/20, and %30/30 agreement rates were 89.6%, 95.3%, and 98.8% for arm-placed sensors and were 85.5%, 93.2%, and 98.1% for abdomen-placed sensors. Across days of wear, glucose concentration ranges, and rates of change, %20/20 agreement rates were accurate within 9% of the overall %20/20. No serious adverse events were reported.

Conclusions:

The G7 CGM provides accurate glucose readings with single-digit MARD with arm or abdomen placement in adults with diabetes.

www red DiabetologNytt